The temperature transducer IM12-
TI02-2TCURTDR-2I-C0/24VDC transmits
temperature-dependent measured values galvanically isolated. The devices are suitable for
operation in zone 2.
The 2-channel device has two inputs for thermocouples acc. to IEC 60584, DIN 43710,
GOST R 8.585-2001, low voltages (-150…
+150 mV), RTDs acc. to IEC 60751, DIN
43760, GOST 6651-94 (2-, 3-wire) and resistors 0…5 kò (2-, 3-wire). Two 0/4 … 20 mA
current outputs are available on the output
side.
The device is parameterized via FDT and
IODD with a PC. The cold junction compensation can be set to internal, external, or to a
constant value. The current outputs can be
set to source or sink 0/4…20 mA. The inputs
can be freely assigned to the outputs. In accordance with the parameterization (E1, E2,
E1 - E2 or E2 - E1), the input signals are provided as a 0/4…20 mA standard current signal.
The devices have a green power LED (Pwr)
and a red LED to indicate internal faults. For
each input circuit there is a yellow and a red
status LED. A fault in the input circuit leads
to a flashing red LED according to NE44, an
internal error to a steady red LED. The fault
current can be adjusted to 21.5
mA.
The device can be used in safety circuits up
to SIL2 (high and low demand according to
IEC 61508) and meets the requirements of
the NE21. It is equipped with removable screw
terminals.
The device is equipped with removable screw
terminals.
■ Input circuits monitored for wire-break and
short-circuit
■ Parameterized via PC
■ Complete galvanic isolation
■ Removable screw terminals
■ ATEX use in Zone 2, cUL
■ SIL 2